| Quadrupole Ion Trap MSGlossaryspace-charge effects: too many ions in the trap distort 
	the electric fields, leading to significantly impaired performance.
	 fundamental rf: a 0.76 MHz potential applied to the 
	ring electrode. The amplitude of this potential determines the mass range of 
	the instrument and is ramped to eject ions.
	 qz: a dimensionless parameter that determines stability of ion 
	trajectories depending upon their mass-to-charge ratio, the size of the ion trap, 
	and the amplitude and frequency of the fundamental rf.
	 qz = 4eV/mr2w2. For the case of the 
	LCQTM traps, r = 1 cm, 
	w/2PI = 0.76 MHz, and V ranges from 0 to 8,500 V0-p.
	 secular frequency: the frequency, dependent upon the 
	qz value, with which an ion oscillates in the trap.
	 resonance: an ac voltage is applied to the endcap 
	and the qz value of an ion of interest is changed until the secular frequency 
	of the ion matches the frequency of the applied ac voltage. When resonance occurs,
	the amplitudes of ion trajectories linearly increase with time. A high-amplitude 
	ac voltage will cause resonance ejection, while a low -amplitude ac voltage will 
	cause resonance excitation
	 ac voltage: also called supplementary or auxiliary potential, 
	is a voltage placed on the endcap electrodes. It can be used to induce resonance 
	excitation and resonance ejection and to isolate ions.
	 tickle voltage: an ac voltage applied to the endcap 
	electrodes during an excitation period. The amplitude of the voltage is generally 
	small so as to enable fragmentation of the ions by collisions with the helium 
	damping gas rather than ejection.
	 bath gas, damping gas, target gas: helium gas in the 
	trapping volume at a pressure of about 1 mtorr serves to cool ion kinetic energies 
	and focus ion trajectories into a tight packet at the center of the trap.  Helium 
	is also used as a collision partner for fragmenting peptide ions via collision-
	induced dissociation.
	 high resolution: an experiment where peaks corresponding 
	to carbon isotopes may be resolved. The mass scan speed of the instrument is 
	reduced, resulting in an increase in the density of data points per unit m/z, 
	thus increasing mass resolution.
 |